CareLink USB Range Testing: Test #2
Setup and Notes
- Date: 2015-09-09
- Pump: Medtronic Minimed 723 v2.4A pump
- Pump Battery: Full
- Pump suspended
- CareLink USB stick plugged directly into RPi2 USB using 6" extension
- Pump and CL stick placed on hardwood floor
- Did not attempt reorientation of pump during test
openaps use pump read_clock
used to test communication;openaps use pump status
andopenaps use pump settings
found to show same range- Successes and failures repeated when pass/fail threshold determined to check for consistency
- Antenna on the CareLink USB stick is on the side of the PCB close to the large "CareLink" text (the "top" side), furthest from the male USB adapter
Results
ID | Stick | Orientation | Pass Distance [ft] | Fail Distance [ft] |
---|---|---|---|---|
01 | Encased | Flat, CL Text Up, Tip Towards Pump | 5.5 | 6.0 |
02 | Encased | Flat, CL Text Down, Tip Towards Pump | 8.0 | 8.5 |
03 | Encased | Flat, CL Text Up, Lower Edge Towards Pump | 9.0 | 9.5 |
04 | Encased | Flat, CL Text Up, Upper Edge Towards Pump | 7.5 | 8.0 |
05 | Encased | Flat, CL Text Down, Upper Edge Towards Pump | 9.0 | 9.5 |
06 | Encased | Flat, CL Text Down, Lower Edge Towards Pump | 9.0 | 9.5 |
07 | Encased | On-Edge, Bottom Towards Pump, CL Text Upside Down | 12.5 | 13.0 |
08 | Encased | On-Edge, Top Towards Pump, CL Text Upside Down | 10.5 | 10.0 |
09 | Encased | On-Edge, Bottom Towards Pump, CL Text Right-Side Up | 11.0 | 11.5 |
10 | Encased | On-Edge, Top Towards Pump, CL Text Right-Side Up | 9.0 | 9.5 |
11 | Naked | Flat, CL Text Up, Tip Towards Pump (see ID 01) | 6.5 | 7.0 |
12 | Naked | Flat, CL Text Down, Upper Edge Towards Pump (see ID 05) | 10.0 | 10.5 |
13 | Naked | On-Edge, Bottom Towards Pump, CL Text Upside Down (see ID 07) | 12.5 | 13.0 |
Analysis
- There is nontrivial directionality when it comes to the CareLink USB stick orientation
- Directional affinity shown here may be related to pump orientation and/or proximity to the hardwood floor, not simply which face is "pointing" towards the pump; this was not explored in these tests
- Removing the plastic case appears to add a marginal amount to the range (0 - 1 ft)