Medical device workflow are rich and complex. This design attempts to factor the domain into a minimal set of components that can be combined to produce useful systems. It's self-consciously naïve, but it should serve as a seed for discussion.
- Any device can be online all the time, and can act as a FHIR server -- possible all on its own, or possibly with the help of a proxy/manager/facade component that "fronts for" the device. (Communication between the device and its manager occurs using any mechanism necessary; but the components below be useful for such internal communication, too.)
(EK: The assumption that a device can act as a server, won't that preclude the use of our resources by less powerful devices? I'm just thinking about smaller worn devices, etc. Maybe just posting results somewhere is the max we can expect from them)
- A device generates snapshot (i.e. point-in-time) reports -- which can consist of observations or alerts.
(EK: Not too sure whether they are both "reports", alerts sound more like inidividually pushed observations rather than reports)
- A device exposes a collection of individual "settings" that can be directly manipulated by authorized clients.
- The FHIR device resources should support:
- The
DEC
(Device Enterprise Communication) profile for (a)periodic physiologic and CLIA waived point-of-care laboratory tests(?) - The
ACM
(Alert Communication Management) profiles for communication of alerts (physiological, technical alarms and advisories)
- What about
PIV
(Point-of-care Infusion Verification),IDCO
(Implantable Device Cardiac Observation),EC
(Event Communication)? DPI
(Device Point-of-care Integration) andMEM
(Medical Equipment Management) are future IHE work an so currently out of scope.
Of interest are the following interactions:
- PCD-01 (Communicate PCD Data)
- PCD-02 (Subscribe to PCD Data)
- PCD-04 (Report Alert)
Note how the Device Observation Reporter fronts for the actual device, although smarter devices may integrate the DOR. Note that the DOR might communicate to the device using proprietary formats (including FHIR). Note also that communication is two-way (have to check out what that means, EK)
In both images, is pushed from the DOR to the DOC, this may be done by:
- the DOR doing a RESTful POST on the DOC acting as a server
- the DOR (as a server) sending a notification to the DOC acting as a client
- the DOR (as a server) returning a result on a (poll) request by the DOC acting as a client
This picture shows the Alert Reporter (the front face for the device) and Alert Manager which received the alerts from the device. Further dissemination of the alert to the human end-user via the Alert Communicator is out of scope.
Device
: as-is in FHIR, describes a specific instance of a device. For example:
{
resourceType: 'Device',
type: {
code: [{
system: 'http://carefusion.com/',
code: 'alaris-pulse-ox'
}]
},
identifier: {
system: 'http://carefusion.com/serial-number/',
value: '75023658923523'
},
location: 'Location/medical_ward_5'
}
DeviceSystem
resource: provides a detailed layout of the components of a device system. Specifically in IHE PCD terms, this describes a tree structure of device, virtualdevices, channels, and metrics. Within this tree, every node has a unique identifier (which can, if desired, be expressed as a PCD-like path). For example:
{
resourceType: 'DeviceSystem',
device: 'Device/pulse_ox_unit_8273513',
virtualDevice: [{
identifier: 'vd1',
channel: [{
identifier: 'vd1/ch1',
metric: [{
identifier: 'vd1/ch1/metric183'
}]
}]
}, {
identifier: 'vd2',
channel: [{
identifier: 'vd2/ch1',
metric: [{
identifier: 'vd2/ch1/metric387'
}]
}]
}]
}
Any given device has a collection of associated DeviceSetting
resources that describe the current state of configurable settings. Each DeviceSetting
represent a single state for a device configuration setting. For example, "the SpO2 alarm threshold" for a specific pulse oximeter. These settings are fully qualified by device and (as needed) path within the virtual device system:
{
resourceType: 'DeviceSetting',
device: 'Device/pulse_ox_unit_8273513',
component: 'vd1/ch1/metric138', # <-- virtualdevice, channel or message id in Device/123
name: {
system: 'http://carefusion.com/alaris-pulse-ox',
code: 'spo2_alarm_threshold'
},
allowedValueRange: {
low: {value: 20, unit '%'},
high:{value: 99, unit'%'}
},
valueQuantity: { # <-- This is the only part a client can change
value: 64,
unit: '%'
}
}
In general, an authorized user can update a DeviceSetting
via HTTP PUT
-- but only the value
portion is allowed to change.
A DeviceObservationReport
captures the complete details of a point-in-time observation report. It contains a set of observations, each associated with a specific component metric. Would consider flattening the structure so the entire hierarchy isn't recapitulated from DeviceSystem
:
{
resource: 'DeviceObservationReport',
device: 'Device/pulse_ox_unit_8273513',
observation: [{
component: 'vd1/ch1/metric138', # <-- virtualdevice, channel or message id in Device/123
observation: 'Observation/123' # <-- e.g. about Patient: "has SpO2 = 82%"
}]
}
A DeviceAlertReport
captures the complete details of a point-in-time device alert. This includes details about what alerts are active and why. A report is explicitly generated, and multiple reports can be generated by the same device in sequence. So a report is not a "singleton"; a whole stream, or feed, of alerts would emerge from a device. It's possible that the leaf-level alert details should be Observation
(Which is how IHE does it in v2) -- I'm not sure.
Importantly: when possible, if an alert is
- triggered by a
DeviceSetting
(e.g. SpO2 threshold underrun), or - can be disabled by a
DeviceSetting
(e.g. "Alarm is buzzing --> false")
then these links should be conveyed here. So right next to the fact that "the alarm is beeping" there's a reference to the settings that allow a client to change that state -- e.g. to change the threshold or simply silence the alarm.
From the IHE PCD TF-1:
The semantics and data types used to represent alert type, alert priority, alert inactivation state 1110 and escalation and de-escalation of priority in the messages of this actor are based on IEC 60601-1-8 definitions. A single source can produce multiple, possibly concurrent, alerts. A single Report Alert transaction can contain at most a single alert.
{
resource: 'DeviceAlertReport',
device: 'Device/pulse_ox_unit_8273513',
alert: [{
component: 'vd1/ch1/metric138',# <-- virtualdevice, channel or message id in Device/123
observation: 'Observation/456', # <-- e.g. about device: "is currently beeping"
wasTriggeredBy: 'DeviceSetting/3897535', # <-- the "SpO2 threshold = 82%" setting
canDisableWith: 'DeviceSetting/9385372' # <-- the "alarm status" setting
}]
}
So a client could GET /DeviceSetting/3897535
, discover the valid settings, and turn the alarm off via HTTP PUT
.